← All writing

Thesis submitted

On submitting and having my PhD thesis confirmed, neuromorphic biosensors, plasma processing, and detecting TDP-43.

A few weeks ago I submitted my PhD thesis. The confirmation came through recently. It's done.

Download thesis

What the thesis was about

The short version: I built neuromorphic electronic devices that can detect specific biomolecules in solution, and demonstrated that they work for TDP-43, a biomarker for neurodegenerative disease.

The longer version requires a bit of background. The devices are electrolyte-gated transistors: three-terminal lateral structures with a liquid channel between the source and drain electrodes, modulated by an applied gate potential. On their own, they behave like ordinary transistors. But when driven with pulsed gate inputs, they produce outputs that resemble postsynaptic signals, spiking currents, paired-pulse depression, spike rate-dependent plasticity. The transistor behaves like a synapse.

The sensing layer is where most of the fabrication work happened. A polymer gate dielectric was modified using plasma immersion ion implantation (PIII), a plasma processing technique that bombards the surface with argon ions at 2.5 kV, introducing free radical covalent binding sites without destroying sub-micron-thick films. Protein enzymes applied to a freshly treated surface bind covalently, and that binding persists for up to 24 hours after processing.

With a functional biofunctionalisation process in hand, I first validated the sensing approach using horseradish peroxidase (HRP), an inexpensive and well-characterised enzyme. Covalently immobilising HRP on the gate layer measurably changed the electric double layer dynamics at the electrodes, which in turn modulated the neuromorphic output characteristics. The device could sense that something was bound.

The more clinically relevant test came with TDP-43. Antibodies against TDP-43 were immobilised on the PIII-modified gate layer. When TDP-43 antigen, present at physiologically realistic concentrations, bound to those antibodies, the postsynaptic output changed in characteristic ways. Critically, using tailored pulse sequences and extracting features like the paired-pulse index, postsynaptic current decay, and cumulative charge, true detection was possible even in the presence of interfering biomolecules.